AE Signal Parameters Explained: Hit, Amplitude, Counts, Energy, RMS

Key points
  • AE signals are often analyzed by summarizing them into features (parameters) rather than storing full waveforms.
  • The basis is calculating amplitude, rise time, duration, counts and energy for each signal that crosses the threshold (a hit).
  • Continuous signals such as leaks are evaluated with RMS or frequency band energy.

AE monitoring data is expressed through parameters such as hits, counts and RMS. This article explains the definition of each parameter and how it is used to judge equipment condition.

Signal waveform on an oscilloscope screen
Photo: Mikasa2025, CC BY 4.0, Wikimedia Commons (cropped)

Why analyze with parameters?

AE signals are sampled at rates of several hundred kHz or more, so continuously storing raw waveforms produces very large data. It is therefore common to record parameters that summarize each signal in a few numbers and to judge equipment condition from their trends.

Threshold Peak amplitude Rise time Duration ● Counts = number of threshold crossings (here: 7)
Figure 1. Main parameters calculated from a single hit

Hit-based parameters

ParameterDefinitionWhat it tells you
HitA signal that crosses the threshold and is recorded as one eventMore hits per unit time indicate increased damage activity
ThresholdThe minimum signal level counted as a hitThe line separating noise from defect signals
Peak amplitudeThe largest signal level within a hit, usually in dBSource strength, distance from the sensor
Rise timeTime from the first threshold crossing to the peak amplitudeIdentifying source type, discriminating noise
DurationTime from the first to the last threshold crossingSignal size, continuity
CountsNumber of threshold crossings within one hitReflects both signal level and duration
EnergyIntegral of the signal magnitude over the hitRelative amount of energy released
Frequency featuresAverage frequency, peak frequency, frequency centroid, etc.Distinguishing sources such as cracks, friction and leaks

Continuous signal parameters

When a signal continues without interruption, as with leaks or friction, it is hard to split into hits. In that case, RMS or ASL (average signal level) — the average signal strength over a period — or the energy in specific frequency bands is used. This is why a rise in RMS is a key indicator in pipeline leak monitoring.

Tips for reading the data

  • Values depend on the threshold. The same signal yields different hit counts, counts and durations at different thresholds, so comparisons require identical settings.
  • Signals attenuate with distance. Amplitude and energy decrease as the distance between source and sensor increases.
  • Trends matter more than absolute values. How hits and energy change over time shows equipment condition better than any single value.
  • Look at several parameters together. Amplitude or counts alone can make it hard to distinguish noise.